This repository analyzes a Co-60 counting experiment with a Geiger-Muller detector. The central question is how much additional absorber thickness is required to keep the net count rate fixed when the source position changes.
This project emphasizes transparent modeling choices: background subtraction, restricted fit ranges, regression inversion, a control test, and generated figures. It is organized as a compact empirical replication package so another reader can rerun the analysis and inspect where each reported number came from.
- Data workflow: count-rate measurements, absorber metadata, and control-test data are turned into regression figures and a compact quantitative summary.
- Methods signal: background subtraction, fit-range restrictions, regression inversion, and an explicit negative-control check.
- Reproducibility signal: a short Python script rebuilds the figures and regression summary from the committed data tables.
The repository is intentionally small: the main figures, regression summary, and control test can be regenerated from the committed tables with one script.
For a target background-subtracted net count rate, how does the required absorber areal density change when the source moves from one slot to another?
- Use background-subtracted count rates for several absorber stacks and source positions.
- Restrict the fit to the gamma-dominated region where beta contributions are negligible.
- Fit net count rate versus areal density for the relevant source slots.
- Invert the fitted lines to compute the equivalent thickness difference
Delta Zat fixed net count rate. - Run a control test to check whether absorber position alone changes the net rate at fixed areal density.
Co-60 decay scheme
Net count rate versus absorber thickness
Equivalent thickness change when moving the source
Control test: absorber position
- Slot 3 fit:
(N - B) = (-0.00857 +/- 0.00167) * Z + (263.00 +/- 12.09) - Slot 4 fit:
(N - B) = (-0.00595 +/- 0.00097) * Z + (177.95 +/- 7.05) - Central mapping over the operating region:
Delta Z ~= 51.22 * (N - B) + 812.66mg/cm^2 - Monte Carlo propagation of the two regression covariance matrices gives coefficient
uncertainties of about
+/- 41.62on the slope and+/- 7081.45 mg/cm^2on the intercept, so the derived mapping should be read as an uncertainty-aware operating estimate rather than a high-precision calibration. - At
N - B = 130 cpm, moving from Slot 4 to Slot 3 requires about7.47e3 +/- 2.13e3 mg/cm^2of additional absorber.
The absorber-position test is a negative control: with total absorber thickness held
fixed, the absorber slot itself should not change the net count rate. A one-way ANOVA
on the repeated measurements gives F = 0.368, p = 0.70, so no detectable
absorber-position effect was found at the fixed areal density used in the control run.
python -m venv .venv
pip install -r requirements.txt
python src/analyze_co60.pyIf make is available:
make allOutputs are written to figures/.
Smoke test:
python -m unittest discover -s testsdata/ Processed attenuation points and raw control-test data
src/ Python analysis script
figures/ Generated plots and regression summary
report/ Technical report
summary/ One-page non-technical summary
assets/ Apparatus photo and decay scheme
- Technical report:
report/report.pdf - One-page summary:
summary/one_page_summary.pdf
Author: Hongyu Wang.
This is a solo analysis and report package: all analysis code, modeling decisions,
figures, and write-up were completed by Hongyu Wang.
Instructor: W. Lippincott, Department of Physics, UC Santa Barbara.